INFICON
======================
VGC094
======================
Software Release Notes
======================

This document describes the software release notes
for the VGC094.



V1.40 [4755]
============
Release Date : 2025-02-21
Filename     : INF_VGC094_V140_4755.S19

New Features
------------
- Setpoint On-Timer resolution changed from 1s to 0.1s.
- Add sensor type menu entry.
- Add GTA/GTB (gauge type) command.
- Update correction curve for PE300DC9 + MAG050/MAG060 combination.
- Add correction curve for CP300C9 + MAG084 combination.
- Add correction curve for CP300C10 + MAG084 combination.
- Add correction curve for CP300T11L + MAG086 combination.
- Add correction curve for PE300DC9 + MAG084 combination.
- Add correction curve for PE300DC9 + MAG050 combination.
- Mnemonics command TIM accepts seconds as optional argument.
- Add sensor ON/OFF/AUTO support to IF500PN process data.
  Requirements: - at least IF500PN firmware version V00.02.01
                - GSDML-V2.40-INFICON_AG
- Show AUTO in "BAUDRATE IFxxx" menu when IF500PN is present.
- BAU command returns 0 (AUTO) when IF500PN is present.
- Add flash memory second source support.
- Add 100Hz anti-aliasing filter to options 1Hz and 0.1Hz.
- Disable debug information displayed when new EEPROM parameters are created.
- Add PE300T11 board support.

Fixed Problems
--------------
- Fix bug where RS485 did not work properly.
- Fix bug where SENSOR -> TYPE was wrong after board switch.
- Enable F.S. COMP. for CP300 boards.
- Mnemonics command SAV non-blocking.
- Show correct status when PE300DC9 sensor is disabled by hardware.
- Prevent BAU command from changing baudrate settings when IF500PN is present.
- Fix bug where "Filter OFF" option did not work properly.
- Fix bug where fatal error could occur during startup depending on boards plugged in.



V1.30
=====
Release Date : 2021-12-07
Filename     : INF_VGC094_V130.S19

Fixed Problems
--------------
- Bugfix: TCP-socket connection failed to send response due to wrong thread priority settings.
- Bugfix: On-Time parameter (SPx commands) is now optional for backward compatibility reasons.
- Bugfix: Prevent, that sensor-ON source can be set to itself (SA1, SA2, SB1 and SB2).
- Bugfix: SPA and SPB commands returned <NAK> when ON source was set
          to PREVIOUS and OFF source to SELF.
- Bugfix: Mnenmonics command BLV returns brand depending application version string now.
- Bugfix: Mnenmonics command SEN returns <NAK> if one of the four parameters a-d is set to 0.
- Bugfix: At IF300P startup the first pressure request command was answered with <NAK>.
- Bugfix: Fixed bug where single bytes received could be lost.
- New:    Set IFxxx baudrate to 115200 automatically if IF500_X_UART is connected.
          IFxxx baudrate cannot be changed manually.
- Bugfix: Fixed bug where TCP socket connection did not work with certain tools.
- Bugfix: Fixed bug where menu did not work properly when watchdog (AUTO) occurred and
          bargraph was enabled.
- Bugfix: Fixed bug where 'Test Menu' message did not pop up as expected during startup.
- Bugfix: Fixed bug where bargraph command DCB did not work properly.
- Bugfix: Fixed bug where even tough "Key Lock" was active, it was
          possible to use shortcuts (e.g. set default, turn sensors ON/OFF, ...).
- New:    Added mnemonics command SME (show me).
- New:    Plug-in boards menu added:
          - Display board identification information
          - Display software version (not available for legacy boards)
          - Display hardware version (not available for legacy boards)
          - Update software via USB memory stick (not available for legacy boards)
- New:    Setpoint ON-Timer added:
          - Menu entry added
          - Added ON-Timer parameter to mnemonics commands SP1, SP2, SP3 and SP4
          - Added ON-Timer functionality to setpoint module.
- New:    Gas-type correction added.
- New:    Major sensor control update:
          - sensor control menu extended
          - added sensor ON options HOTSTART + xx, PREVIOUS [+ xx]
- Bugfix: FATAL ERROR if firmware image is downloaded to a device with bootloader V1.56 and older.
          Device ID was introduced in bootloader V1.57 and application fails to read value
          if parameter does not exist.
- Bugfix: Response buffer was never freed when send failed.
- Bugfix: Ethernet settings could not be changed after "LOAD DEFAULTS".
- New:    Leakage current compensation added to Penning type sensors.
          - Added entry to menu SENSOR
          - Added commands CA1, CA2, CB1 and CB2
- Bugfix: Fixed a bug where device might reboots when reading sensor control thresholds using IF500PN.
- Bugfix: Fixed bug where fatal error could occur during startup depending on boards plugged in.
- New:    Disabled debug information displayed when new EEPROM parameters are created.
- Bugfix: Fixed bug where initial boot failed.



V1.20 - INITIAL RELEASE
=====
Release Date : 2021-03-15
Filename     : INF_VGC094_V120.S19

Known Problems
--------------
- 