INFICON
======================
VGC501
VGC502
VGC503
======================
Software Release Notes
======================

This document describes the software release notes
for the VGC501, VGC502 and VGC503.



V1.08
=====
Release Date : 2024-04-24
Filename     : INF_VGC50x_V108.S19

New Features
------------
- Support for BAG5xx, BPG5xx and BCG552 gauges
- new Mnemonic interface commands
  - 'GIM' GaugeType
    The new command 'GIM' (Gauge Identification Mode)
    makes it possible to switch the identification of
    each measuring channel from 'AUTO' to a fixed gauge
    type.
    On the one hand, this can prevent errors due to
    incorrect identification in critical applications
    with large sources of interference.
    On the other hand, gauges without an identification
    resistor and a different output characteristic can
    now also be operated on the VGC50x.
    The default setting of the 'Gauge Identification
    Mode' is 0='AUTO'.
  - 'GFx' Gauge x Formula
    With the additional commands 'GFx' (Gauge x Formula,
    per measuring channel) and using three parameters
    (a,b,c), almost any conversion from voltage to
    pressure can be carried out.
    Two formulas are available for either logarithmic
    or linear characteristics.
    - p = 10^((U - a) / b + c) (logarithmic characteristic)
    - p = U * a + b (linear characteristic)
    The default setting of the 'Gauge x Formula' is
    '6.143,1.286,0'. These values correspond to the
    logarithmic characteristic of the PSG500 gauge.
- Extended measuring range for Pirani gauges
  down to 1.1E-4 mbar (previously 5.0E-4 mbar)



V1.07
=====
Release Date : 2021-03-09
Filename     : INF_VGC50x_V107.S19

Fixed Problems
--------------
- Support for new displays (production line changed)
- Stability improvements for TCP socket (port 8000)
  The socket connection is reopened and released for
  new connections after an error occurred.
- Non-volatile saving of parameters 'EMISSION' and
  'UNIT' to the gauge didn't work - corrected



V1.06
=====
Release Date : 2018-02-23
Filename     : INF_VGC50x_V106.S19

New Features
------------
- The heat up status of the digital CDG is now shown
  on the display with the heating wire symbol
- New value 'p = f (0.5h)' for parameter BARGRAPH
  for long time (> 2 days) trend display
- New value 'IDENT' for parameter BARGRAPH
  displays the sensor identification of the active
  measuring channel permanently
- Added floating point representation for logarithmic
  gauges when the FORMAT setting is 'X.X' and
  UNIT 'PASCAL'
- New AO-MODE's : PRL10K / PRL1Q (nonlinear, Edwards)

Fixed Problems
--------------
- With the key-lock function, sensor on/off by
  keyboard is now also locked
- COR factor does now also affect SP setting range
- Fixed bug - an additional <CR><LF> sent after the
  <ENQ> character could reboot the device
- A sensor fault now always switches off the
  controlled sensor
- Fixed bug - when increasing the pressure of a HPG
  gauge from the Ioni to the Pirani range, a pressure
  in the E-6 range is displayed for a moment
- PEG100 does not switch on (~1s) anymore at power on
  of the VGC501

Known Problems
--------------
- 



V1.05 (not released for production !)
=====
Release Date : 2017-01-26
Filename     : INF_VGC50x_V105.S19

New Features
------------
- New parameter 'CALIB' in the Test Parameter section
  to set a re-calibration date.
  A warning is displayed when the date is reached.

Fixed Problems
--------------
- 

Known Problems
--------------
- 



V1.04
=====
Release Date : 2016-11-02
Filename     : INF_VGC50x_V104.S19

New Features
------------
- new AO-MODE's : PRM10K (nonlinear, Edwards)
                  IMR110 (p = 10^[U/2 - 6])
                  IMR120 (p = 10^[U/2 - 8])
                  IMR310 (p = 10^[U*6/10 - 6])
                  IMR320 (p = 10^[U*7/10 - 9])
- New option to specify a unit of measurement when
  editing parameter setups :
  If saving parameter setups to a USB-stick, the
  values for setpoint/threshold values and offset
  are always stored in the unit mBar/hPa.
  If restoring parameter setups from a USB-stick
  and no unit has been specified in the file for the
  setpoint/threshold values and the offset, the value
  is read in the unit mBar/hPa.
  Otherwise, one of the measurement units 'MBAR',
  'HPASCAL', 'TORR', 'PASCAL' or 'MICRON' must be
  explicitly entered in capital letters and with a
  single space, eg. '5.00-4 TORR' or '0.0002 PASCAL'.

Fixed Problems
--------------
- 

Known Problems
--------------
- 



V1.03
=====
Release Date : 2016-10-03
Filename     : INF_VGC50x_V103.S19

New Features
------------
- AO-MODE "LOG C1" - now PCG gauges also supported

Fixed Problems
--------------
- "CPR" Mnemonic command - Measurement value had only
  2 decimal places instead of 4
- "PR1", "PR2", "PR3" and "PRX" Mnemonic commands -
  Number format for linear gauges was sx.xx00Esxx
  instead of sx.xxxxEsxx.
  Number format for logarithmic gauges was
  sx.xxxxEsxx instead of sx.xx00Esxx.
- HV-Pins at the CONTROL connector now control
  correctly the sensor when powering the VGC50x
- FSR "DIxxxx" in OLV compatibility mode is no longer
  reset when powering up
- Setpoint settings for SP-H are now adopted
  correctly for linear sensors at startup. Mistakenly
  an adjustment of 10% was made at the start before.
..... and some more small improvements

Known Problems
--------------
- 



V1.02
=====
Release Date : 2015-12-10
Filename     : INF_VGC50x_V102.S19

New Features
------------
- New screensaver option "DR" (darkroom) to switch
  off backlight completely after 1 minute of
  inactivity
- "DCS" Mnemonic command expanded for new darkroom
  option (value 6)
- Test-Menu - Additional method to jump to the test
  menu by pressing the menu-button for more than 5
  seconds when the parameter group menu shows '<'

Fixed Problems
--------------
- RNG-EXT parameter is displayed now as soon as at
  least one PSG or PCG gauge is connected
- User interface - Easily adapted translations
..... and some more small improvements

Known Problems
--------------
- "CPR" Mnemonic command - Measurement value has only
  2 decimal places instead of 4
- "PR1", "PR2", "PR3" and "PRX" Mnemonic commands -
  Number format for linear gauges is sx.xx00Esxx
  instead of sx.xxxxEsxx.
  Number format for logarithmic gauges is sx.xxxxEsxx
  instead of sx.xx00Esxx.



V1.01
=====
Release Date : 2015-08-13
Filename     : INF_VGC50x_V101.S19

Fixed Problems
--------------
- Operating language was not changeable in V1.00,
  has been corrected

Known Problems
--------------
- 



V1.00 - INITIAL RELEASE
=====
Release Date : 2015-07-07
Filename     : INF_VGC50x_V100.S19

Known Problems
--------------
- 