$ Electronic Data Sheet for the TTR211S-D
$ Leybold Vakuum GmbH
$
[File]
DescText = " Leybold Vakuum TTR21XS-D EDS File - TTR2";
CreateDate = 08-08-97;
CreateTime = 12:00:00;
ModDate = 10-29-2003;
ModTime = 12:00:00;
Revision = 1.02;
[Device]
VendCode = 144; $ Vendor Code
VendName = "Leybold Vakuum";
ProdType = 0; $ Product Type
ProdTypeStr = "Generic";
ProdCode = 2; $ Product Code
MajRev = 1; $ Major Rev
MinRev = 0; $ Minor Rev
ProdName = "TTR211S-D";
Catalog = "89651";
[IO_Info]
Default = 0x0001; $ Poll, BitStrobe, COS
PollInfo = 0x0007,
1, $ Default: Producing 3
1; $ Default: Consuming 1
StrobeInfo = 0x0007,
3, $ Default: Producing 5
1; $ Default: Consuming 1
COSInfo = 0X0007,
1, $ Default: Producing 3
1; $ Default: Consuming 1
$ Input Assemblies
Input1 = 6, $ bytes to be transfered
0, $ significant bits
0x0007, $ Compatible I/O type mask
"Input Assembly 3",
6, "20 04 24 03 30 03", $ Connection Path Size/Path
"Status, Except, Press"; $ Help String
Input2 = 14, $ bytes to be transfered
0, $ significant bits
0x0007, $ Compatible I/O type mask
"Input Assembly 4",
6, "20 04 24 04 30 03", $ Connection Path Size/Path
"Status, Except, Press, TripPoint"; $ Help String
Input3 = 4, $ bytes to be transfered
0, $ significant bits
0x0007, $ Compatible I/O type mask
"Input Assembly 5",
6, "20 04 24 05 30 03", $ Connection Path Size/Path
"Pressure"; $ Help String
Input4 = 1, $ bytes to be transfered
0, $ significant bits
0x0007, $ Compatible I/O type mask
"Input Assembly 6",
6, "20 04 24 06 30 03", $ Connection Path Size/Path
"Exception"; $ Help String
Output1 = 1, $ bytes to be transfered
0, $ significant bits
0x0007, $ Compatible I/O type mask
"Output Assembly 1",
6, "20 04 24 01 30 03", $ Connection Path Size/Path
"Trip Source"; $ Help String
Output2 = 9, $ bytes to be transfered
0, $ significant bits
0x0007, $ Compatible I/O type mask
"Output Assembly 2",
6, "20 04 24 02 30 03", $ Connection Path Size/Path
"Trip Source, Trip Point"; $ Help String
[ParamClass]
MaxInst=11;
Descriptor=0;
CfgAssembly=0;
[Params]
$ SENSOR PRESSURE OBJECT
$ ****************************************************************************
$ * Parameter Instance #1: Sensor State
$ ****************************************************************************
Param1=
0, $ parameter value slot
6, "20 67 24 01 30 03", $ link size and path
0x0032, $ descriptor (RO, Mon, Enum)
0xC1, 1, $ BOOL, 1 bytes
"Sensor State", $ parameter name
"", $ units string
"Gauge ON == Normal Operation ",
0, 0x1, 0, $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 0 ; $ scaling links and precision
$ TRANSFORM PRESSURE OBJECT
$ ****************************************************************************
$ * Parameter Instance #2: Pressure Value
$ ****************************************************************************
Param2=
0, $ parameter value slot
6, "20 68 24 01 30 01", $ link size and path
0x0030, $ descriptor (RO, Mon, Enum)
0xca, 4, $ Real, 1 bytes
"Pressure Value", $ parameter name
"", $ units string
"The pressure value is displayed in the actual unit mbar, Torr or Pascal",
0.0005, 100000, 0.0005, $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 4; $ scaling links and precision
$ ****************************************************************************
$ * Parameter Instance #3: Pressure Unit
$ ****************************************************************************
Param3=
0, $ parameter value slot
6, "20 68 24 01 30 03", $ link size and path
0x0022, $ descriptor (Mon, Enum)
0xc6, 1, $ USINT, 1 bytes
"Pressure Unit", $ parameter name
"", $ units string
"mbar , Torr , Pascal",
0,2,2, $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 0 ; $ scaling links and precision
$ DISCRETE OUTPUT POINT
$ ****************************************************************************
$ * Parameter Instance #4: TP State
$ ****************************************************************************
Param4=
0, $ parameter value slot
6, "20 69 24 01 30 03", $ link size and path
0x0032, $ descriptor (Mon, Enum)
0xC1, 1, $ BOOL, 1 bytes
"Trip Point State", $ parameter name
"", $ units string
"0 relais deactivated, 1 relais activated ",
0, 0x1, 0, $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 0 ; $ scaling links and precision
$ ****************************************************************************
$ * Parameter Instance #5: TP Control Source
$ ****************************************************************************
Param5=
0, $ parameter value slot
6, "20 69 24 01 30 69", $ link size and path
0x0022, $ descriptor (Mon, Enum)
0xC1, 1, $ BOOL, 1 bytes
"Trip Point Control Source", $ parameter name
"", $ units string
"You can define whether the trip point is installed by DeviceNet or by the potentiometer of the ITR100-D",
0, 0x1, 1, $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 0 ; $ scaling links and precision
$ ****************************************************************************
$ * Parameter Instance #6: TP Upper Value
$ ****************************************************************************
Param6=
0, $ parameter value slot
6, "20 69 24 01 30 67", $ link size and path
0x0020, $ descriptor (RO, Mon, Enum)
0xca, 4, $ Real, 1 bytes
"Trip Point Upper Value", $ parameter name
"", $ units string
"Only active if Trip Point Control Source == DeviceNet",
0.001, 83300, 0.001 , $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 4; $ scaling links and precision
$ ****************************************************************************
$ * Parameter Instance #7: TP Lower Value
$ ****************************************************************************
Param7=
0, $ parameter value slot
6, "20 69 24 01 30 68", $ link size and path
0x0020, $ descriptor ( Mon, Enum)
0xca, 4, $ Real, 1 bytes
"Trip Point Lower Value", $ parameter name
"", $ units string
"Only active if Trip Point Control Source == DeviceNet",
0.001, 83300, 0.001 , $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 4; $ scaling links and precision
$ ANALOG OUTPUT POINT
$ ****************************************************************************
$ * Parameter Instance #8: Analog Output Mode
$ ****************************************************************************
Param8=
0, $ parameter value slot
6, "20 70 24 01 30 65", $ link size and path
0x0032, $ descriptor (Mon, Enum)
0xc6, 1, $ USINT, 1 bytes
"Analog Output Mode", $ parameter name
"", $ units string
"Log: Ua=(log(p/mbar)*1.286 +6.143 " ,
0, 0, 0, $ min, max, default
1, 1, 1, 0, $ mult, div, base, offset scaling
0, 0, 0, 0, 0 ; $ scaling links and precision
[Groups]
$ Group1="SPO",1;
$ Group2="TPO",2,3;
$ Group3="DOP",4,5,6,7,8,9;
$ Group4="AOP",10;
[EnumPar]
$ ****************************************************************************
$ * Parameter Instance #1: Sensor State
$ ****************************************************************************
Param1=
"FAIL",
"ON";
$ ****************************************************************************
$ * Parameter Instance #3: Pressure Unit
$ ****************************************************************************
Param3=
"mbar",
"torr",
"Pascal";
$ ****************************************************************************
$ * Parameter Instance #4: TP State
$ ****************************************************************************
Param4=
"OFF",
"ON";
$ ****************************************************************************
$ * Parameter Instance #5: TP Control Source
$ ****************************************************************************
Param5=
"Potentiometer",
"DeviceNet";
$ ****************************************************************************
$ * Parameter Instance #8: Analog Output Mode
$ ****************************************************************************
Param8=
"log";